HOME  ㅣ  CONTACT US  ㅣ  JAPANESE  ㅣ  ADMIN
COMPANY PRODUCTS TECHNICAL INFORMATION SUPPORT NEWS
X-ray Dose Meters

Quart dose meters for diagnostic devices and CT have been manufactured for the high convenience of the user. 
QUART's expertise and knowledge cultivated over 30 years have reduced complicated pre-settings.
Since the necessary evaluation items for QA can be required with a single irradiation, the burden of QA / QC work can be reduced. Also, it sticks to the design, is easy to check visually, and it is easy to carry and manage due to compact and lightweight design.


Dose meters for advanced diagnostic device


■ QUART didoNEO R

This is a multi-modality dosimeter that can be used for general X-ray, fluoroscopy, dental, etc.
It is the most compact design of the palm size, and possible to measure high-precision dosimetry.
Only one exposure is required to collect major evaluation items (Dose, Dose rate,kV, Exposure time, Waveform, Total filtration, HVL).

- The most compact multi-parameter sensor
- The smallest multi-functional base unit in similar dose meter
- Can see waveform preview on the unit’s display (also possible by PC)
- Up to 20.000 exposures can be stored in full.
- A couple of optional extras are available for functional enhancements of the system.
- Technical upgrades will enable additional functions in the future.

 

DOSE Range
Uncertainty
Mode
0.25 µGy ~ 999 mGy
±5 %
Added filtration / No added filtration
Automatic Compensation for general x-ray radiation qualities
DWP,DLP Range
Uncertainty
Mode
Feature
0.25 µGy ~ 999 mGy
±5 %
Added filtration / No added filtration
Automatic Compensation for panorama and dent x-ray radiation qualities
DOSE RATE
Range
Uncertainty
Mode
Feature
0.3 µGy/s ~ 999 mGy/s
±5 %
Real time / Maximum / Average
Period dose rate after first half of exposure (for dose calculation according DIN 6868-151 for dental OPGs)
kV Range
Uncertainty
Mode
50 ~ 150 kV (R)
±5 % at filtration of 2-8 mm Al
Maximum(kVp)/ Effective kV(PPV)
TIME Range
Uncertainty
Mode
0.2 ms ~ 300 s
0.1% or ±0.2ms > 1 µGy/s、 0.1 % or ±250ms < 1 µGy/s
Radiation time / Exposure time
PULSE Range
Uncertainty
1 ~ 65.000
±1 Pulse
TF Range
Uncertainty
Width
0.8 ~ 9 mm Al
±8 %
60 ~ 100 kVp、0.8 ~ 8.5mm Al
HVL Range
Uncertainty
Width
0.1 ~ 11 mm Al
±8 %
0.1 ~ 12.5 mm Al


■ QUART didoEASY R (General X-ray, Fluoroscopy)

■ QUART didoEASY M (Mammo)

■ QUART didoEASY MR (Both General X-ray and Mammo)

QUART didoEASY can be used for almost all modalities such as general x-ray, mammography, fluoroscopy, DSA, dental, and dental CT. There is no need for presetting, and high-precision dosimetry is easily possible.
Ideal for measuring parameters required for periodic service and quality assurance of X-ray equipment. Also, this can directly measure DWP(dose-wide product).

- Measures dose, dose rate and exposure time
- Small and lightweight semiconductor type sensor
- Simple and highly accurate dosimeter
- Easy operation without presetting (Positioning → irradiation → measured value display)
- Automatic compensation for ALL beam qualities
- No further corrections or compensations required
- Direct dose-width product (DWP) measurement at dental OPGs

 

APPLICATION Range 40 to 160 kV(didoEASY R)
25 to 40 kV(didoEASY M)
25 to 160 kV(didoEASY MR)
SPECIAL FEATURE   Automatic compensation for main radiation qualities and beam filtration on:
・Dental
・RAD and Fluoro
・Mammography (didoEASY M and MR only) No pre-setting procedure required
No additional corrections required
CUSTOMISATION SERVICE   Customizable beam quality calibration available on request!
DOSE Range Resolution Uncertainty 0.2 μGy ~ 999 Gy
0.01 μGy
< 5 %
DOSE-WIDTH PRODUCT
(DWP)
Range Resolution Uncertainty 0.2 μGy*cm ~ 999 Gy*cm (50 ~ 150 kV)
0.01 μGy*cm
< 5 %
DOSE RATE Range Resolution
Uncertainty
Modes (3):
0.25 μGy/s ~ 999 mGy/s
0.01 μGy/s
< 5 %
Real time display / Maximum rate / Total average
TIME Range Resolution Uncertainty Time Mode 0.5 ms ~ 300 s
0.1 ms
< 0.5 % (± 0.5 ms)
Radiation time

CONFIGURATIONS
• dido EASY+:DOSE, DWP, RATE , , , E, exposure Time
• dido EASY:DOSE , DWP , RATE , Exposure Time



Precision survey meter


■ QUART didoSVM

The QUART didoSVM medical survey meter is designed to detect beta, gamma and x-ray sources of very low intensity. It can sense up to nSv, and you can use it for a wide range of applications. Since the semiconductor chamber is extremely small and lightweight, by attaching an extension rod (optional), it is possible to easily measure areas difficult to measure such as under the ceiling. Furthermore, unlike the conventional ionization chamber type, one of the features is the overwhelming response speed (less than 1 second) for measuring air kerma rate and air kerma from the sources of X ray, β ray, and γ ray. The same measurement result can be obtained no matter who measures it, and while the measurement is being executed in order to track the characteristics of the dose rate. The dose rate of the meter display continues to be updated.

- compact and light-weight semiconductor detector
- light weight base unit
- fast response time to radiation
- reproducible measurement results
- accurate detection of signals against background noise
- detects radiation from leakage, scatter beams and pinholes
- detector and base unit connected magnetically for one-hand use
- detector mountable on tripod or a telescopic extension for measurements in heights up to approx. 3.5 meters above ground
- backlit display to assure readings in dark environments
- powered by rechargeable battery
- approximately 80 hours of continuous use
- recharging duration 3 - 4 hours
- low battery warning

 

OPERATING RANGE   15 keV ~ 2 MeV (Auto-Ranging)
Above 20 keV for gamma and x-rays Above 1 MeV for beta radiation
DOSE Range Resolution 3 nSv ~ 10 Sv
0.1 nSv
DOSE RATE Range Resolution 0.1 μSv/h ~ 2 Sv/h
0.1 μSv/h
TIME Range 0.5 s ~ 15 min
UNCERTAINTY   < 10 % (N-20 to N-300)
< 15 % (Cs-137 & Co-60)
RESPONSE   < 1.0 s (for the full dynamic range)
DISPLAY   Digital numeric value refreshed every second, Analog Bar Graph in three divisions according pre-defined danger levels*
・3.2 µSv/h
・10 µSv/h
・3 mSv/h
・or: factory pre-set as per user requirement
AUDIO OUTPUT   Signal frequency dependent on danger level

Danger Levels in accordance with German Labour Protection regulations:
• Radiation Protection Act
• X-Ray Appliance Act



Meter for CT


■ QUART didoCT

QUART didoCT is a semiconductor type dosimeter for CT, it is possible to easily measure high-precision DWP (dose-wide product) without presetting. By using a semiconductor in a chamber, unlike the conventional ionization chamber type, it does not suffer from changes due to ambient temperature or atmospheric pressure, so changing the settings is not necessary. Also, while the measurement is being executed in order to track the characteristics of the dose rate, the dose rate of the meter display continues to be updated. There is HVL for CT as an option.

 

DOSE-WIDTH PRODUCT (DWP) Range
Uncertainty
0.2 μGy*cm ~ 999 mGy*cm
±5 %
DWP Rate Range
Modes (3)
Uncertainty
0.25 μGy/s*cm ~ 999 mGy/s*cm
Real time / Total average / Maximum Rate
±5 %
CT-HVL

Range
Exposure Conditions

Uncertainty

3.0 ~ 9.0mm Al(HVL output)
2.8 ~ 29mm Al equivalent and corresponding range
Of 70 ~ 150 kVp
±8 %
Exposure Time Range
Mode
Uncertainty
0.5 ms ~ 300 s
Duration of full exposure
± 0.5 ms
Weight Base unit
Detector unit
180 g
120 g (including cable)
Size Base unit
Detector unit
17 x 7 x 4.5 cm (LxWxH)
16 x 1.3 cm (Lx Diameter)
Active length up to 100 mm (marked)


■ CTDI measurement Phantom

The 3-part CTDI phantom can be used with any CT system and may be used to image and monitor the adult head and body, as well as pediatric dose requirements.
The phantom set consists of a group of head, body, and pediatric PMMA sections with five probe holes in each section.
They are located in the phantom bodies according to the standard requirements. Acrylic rods are provided to seal the unused holes.
2-Part CTDI Dosimetry Phantom (Adult Body and Head) is also available.

 

SIZE Adult section-diameter
Pediatric Body / Adult Head
Pediatric Head
Diameter: 320mm, Length: 150mm
Diameter: 160mm, Length: 150mm
Diameter: 100mm, Length: 150mm
Measurement hole size   Diameter:13.1mm
(1 center, 4 peripheral,10 mm from edges; holes to be sealed with adjoining acrylic rods when not in use).