HOME  ㅣ  CONTACT US  ㅣ  JAPANESE  ㅣ  ADMIN
COMPANY PRODUCTS TECHNICAL INFORMATION SUPPORT NEWS
Various Meters

Quart X-ray field meter and the high-precision kV / mass meters are designed to be able to measure with simple operation without trouble of the inspector.
It can be used without troublesome and complicated presetting, and very high accuracy measurement value can be acquired.


X-ray field & Fan beam meter


■ QUART nonius

QUART nonius is a sophisticated measuring instrument for easily checking the geometrical characteristics of X-ray irradiation field and X-ray. It can be used in both analog and digital systems and can be used for analysis of X-ray fan beam characteristics of CT and dental equipment. Measurement accuracy is very high, and its tolerance is only 0.1 mm. Because of its excellent precision and resolution properties, enhanced product research has validated the nonius to be suitable for QA application in Radiation Therapy. Measurement results are transferred to the PC in real time, and the data is visualized in an easy-to-understand manner by the nonius software.
It can be connected to any PC as long as it uses Windows OS. Also, it can measure the position and width of the X-ray of the fan beam, as well as dose.

- optional QUART bridge holder available as an accessory
- nonius is equipped with a USB cable and comes with the associated software & a compact carrying case.

[How to use] Very easy to use!
① Connect the device via USB to a Laptop or Tablet PC
(Windows OS required).
② Position the head unit at the respective position.
③ Use light field or a reference point for alignment.
④ Trigger the QA/QC exposure.
⑤ Immediately evaluate the results.



Accuracy / resolution ±0.1 mm
Exposure Threshold Dose ≥ 200 μGy
Dose Rate ≥ 20 μGy/s
Minimum Exposure Variable; depends on application
Measurement Method Open measurement – no added filtration
Connectivity Standard USB (2.0)
Plug and Play Component
Operation Temperature Range
Storage Temperature Range
15 ~ 40 °C
0 ~ 50 °C
System Requirements Operating System Pentium III、128 Mb RAM、USBx1
Windows 10,7, Vista, XP
Sensor Area 40 mm Length (16 Active Sensor Elements) Radiopaque center marker (Visible in test exposure) Improved sections for light field and laser markings
Weight 190 g (without USB Cable)
Size of Head Unit 55 x 75 x 15 mm (W x L x H)


The high-precision kV / mAs meters


■ QUART didoMAS

QUART didoMAS is a high precision meter that can be used with almost all X-ray equipment from general imaging to mammography. The measurement range is set automatically, and presetting for measurement time and parameters is unnecessary.
A customized cable for the connection and detector head and circuit are included in the set.

- The mA is refreshed and displayed every second.
- One charge, 80 hours of continuous use.
- Recharging the meter until full takes only between 3 – 4 hours
- Low battery warning

 

mAs Range
Uncertainty
0.001 – 9999 mAs
+/- 0.01 mAs or 1 %
mA Range
Modes (3)
Uncertainty
0.1 – 999 mA
・Refresh (4 times per second) during exposure
・Average current
・Maximum
+/- 0.1 mA or 0.5 %
Exposure Time Range
Modes
Uncertainty
1.0 ms – 300 s
Duration of current flow
± 1.0 ms (for rectangular signal)
WEIGHT Base unit
Detector unit
180 g
120 g including cable
SIZE Base unit
Detector unit
170 x 70 x 45 mm (LxWxH)
47 x 30 x 40 mm (LxWxH)


■ MAS1

MAS 1 can measure electric current from various X-rays and display integrated type ammeter. Since this meter has an AC / DC automatic detection function, resetting manually is not needed. Also, since the display of the meter shows all the information at one time, one irradiation is enough.

- measures electric current from any x-ray: single phase / full wave / multiphase /DC
- use for medical or dental x-rays, digital x-ray
- battery operated, built in measurement of battery
- displays mA, mAs and exposure time for each reading

 

APPLICATION
(mA)
200 mA Full Scale: 5 – 200 mA / 0.1 mA Resolution
2 A Full Scale: 1 mA Resolution
ACCURACY
(mA/mAs)
1 % +/- 2 mA on all ranges
CONNECTION Tip Jacks, Test leads with two alligator clips
1 meter cable with two alligator clips
DISPLAY 5.5 mm, (0.22“) Liquid Cr ystal, 2 lines,
12 Character Alphanumeric
CONTROLS ON / OFF / Range Switch
POWER 9 volt battery accessible
(BATTERY LIFE 100 hours continuous)
Low battery indicator (After over one year of normal use)
OP CONDITIONS +10 ~ 40 °C
SIZE/WEIGHT 80 X 147 X 40 mm / 250g


■ MAK1/MAK1L

MAK1 and MAK1L are multi voltmeters that can measure current as well as voltage.
Either direct current or alternating current x-rays can be used, no pre-setting is required.
It is battery-operated and can be used in various places.

- optimized for dental x-rays, also works on radiographic and fluoroscopic x-rays.
- solid-state digital design
- direct measurement of peak kV from the x-ray head
- indicates x-ray waveform type (half-wave, full-wave or DC / 3-phase)

 

kV Range 45 ~ 125 kV (MAK1)
40 ~ 115 kV (MAK1L)
ACCURACY
kVp
 
 
 
Time
mAs
2 % +/- 1 kV, at 25 to 100 mA (MAK1)
or 10 to 50 mA (MAK1L) both
18 to 42 cm from head, for Tungsten target x-ray
tube with 1.5 mm Aluminum equivalent filtration.
1 % +/- 2 ms (1/5 to 2 sec)
Specifications same as MAS1 (previous page)
MIN. EXPO TIME   100 millisec-High Speed Mode
200 millisec-High Resolution Mode
MIN. CURRENT   7 mA (MAK1) or 5 mA (MAK1L) at 50 kV, 10 cm from x-ray
SIZE / WEIGHT   150 x 120 x 58.5 mm / 0.7Kg


■ K2/K2L

K2 and K2L are voltmeters for tungsten generators.
It can be used with various X-ray equipment such as single phase / full wave / multiphase / direct current.
It can be directed measurement of peak kV from the x-ray head.
K2 is for normal range.
K2L is suitable for sensitive range.
Irradiation time is measured at each second, it automatically detects AC or DC and automatically resets after measurement.

- optimized for dental x-rays, also works on radiographic and fluoroscopic x-rays.
- solid-state digital design
- indicates x-ray waveform type (half-wave, full-wave or DC / 3-phase)

 

kV Range 45 ~ 125kV (K2)
40 ~ 115kV (K2L)
ACCURACY
kVp
 
 
 
Time
2 % +/- 1 kV, at 25 to 100 mA (K2)
or 10 to 50 mA (K2/K2L)
18 to 42 cm from head, for Tungsten target x-ray
tube with 1.5 mm Aluminum equivalent filtration.
1 % +/- 2 ms (1/5 to 2 sec)
MIN. EXPO TIME   100 ms - High Speed Mode
200 ms - High Resolution Mode
DISPLAY   10.2 mm (0.4“), Liquid Crystal, 8 Character Alphanumeric
CONTROLS   ON/OFF  SWITCH、 MODE SWITCH
POWER   4 AA batteries accessible
(Battery life 48 hours continuous)
Low batt. indicator (typically 9 months of normal use)
Back